Valleylab Force 2 Electrosurgical Generator Repair Evaluation

$50.00

Full system diagnosis and repair evaluation. We will contact you with repair price prior to completing the repair.

Valleylab Force 2 Electrosurgical Generator
The Valleylab Force 2 Electrosurgical Generator provides cutting and coagulation in monopolar or bipolar outputs. Force 2 has audible and visual alarms, including REM™ Fault Indicator. The REM (Return Electrode Monitoring) is standard. 300 watts is available for output in the cut setting. A monopolar foot switch is included, a bipolar foot switch is optional as is a rolling cart.

Valleylab Force 2 Electrosurgical Generator Features
BLENDED CUT MODES – Provide flexibility through varying degrees of hemostasis. Choose one of three preset modes.

SIMULTANEOUS INDEPENDENT COAGULATION – Permits two surgeons to fulgurate from a single generator, for added convenience and efficacy. Accessories are activated only when keyed, reducing the potential for injury caused by inadvertent activation.

LOW VOLTAGE COAGULATION – Ensures controlled, precise desiccation with less destruction of peripheral tissue, making the Force 2 generator ideal for laparoscopic procedures.

REM SAFETY – Is guaranteed with Valleylab’s patented adaptive REM system, which continually monitors patient impedance levels. If a fault in the patient/return electrode contact is detected, the REM system automatically deactivates the generator virtually eliminating the risk of burns under the return electrode. Valleylab REM safety has been proven in more than 100 million surgical procedures worldwide.

Valleylab Force 2 Electrosurgical Generator Specifications
Dimensions: 13″ W x 21″ D x 8″ H
Weight: 23 lbs
Input Power
Nominal input line voltage is 120 VAC with a full range of 85-135 VAC
Frequency line varies from 45 to 65 Hz
Maximum Consumption: 300 W
Output Waveform
Cut – 500 kHz Sinusoid
Blend 1 – 500 kHz bursts of sinusiod at 50% Duty Cycle, at 31 kHz Recurrence
Blend 2 – 500 kHz bursts of sinusiod at 37.5% Duty Cycle, at 31 kHz Recurrence
Blend 3 – 500 kHz bursts of sinusiod at 25% Duty Cycle, at 31 kHz Recurrence
Coag – 500 kHz damped sinusiodal burst with a repetition frequency of 31 kHz
Low Volt Coag: 500 kHz bursts of sinusiod at 25% duty cycle, recurring at 31 kHz
Bipolar – 500 kHz sinusiod, Unmodulated
Output Characteristics
Cut: Max P-P 3300, Rate Load 300, Nominal Power 300, Crest Factor 1.9 @ 100W
Blend 1: Max P-P 3500, Rate Load 300, Nominal Power 250, Crest Factor 3.3 @ 100W
Blend 2: Max P-P 3700, Rate Load 300, Nominal Power 200, Crest Factor 4.0 @ 100W
Blend 3: Max P-P 4000, Rate Load 300, Nominal Power 150, Crest Factor 4.8 @ 100W
Coag: Max P-P 7000, Rate Load 300, Nominal Power 120, Crest Factor 9.0 @ 50W
Low Volt Coag: Max P-P 4000, Rate Load 300, Nominal Power 99, Crest Factor 4.8 @ 100W
Bipolar: Max P-P 800, Rate Load 100, Nominal Power 70, Crest Factor 2.0 @ 40W
Low Frequency Leakage (50/60 Hz)
Normal polarity, intact chassis ground: less than 10 mA
Normal polarity, ground open: less than 100 mA
Reverse polarity, ground open: less than 100 mA
Sink current, 140 V applied, all inputs: less than 150 mA
High Frequency Leakage
Less than 150 mA rms
REM™ Contact Quality Monitor
Measurement Frequency: 140 kHz
Measurement Current: Max 3 mA
Acceptable Resistance Ranges
REM pad 5-135 ohms
Non-REM pad less than 20 ohms
Cooling: Convection, no fan
Audio Volume: The mode indicator tones are adjustable to a minimum level of 45 dB at 1 meter. The alarm tones are not adjustable and are set at 65 dB minimum at 1 meter.